Features

No calibration is required

It requires 0,6 µL blood. It’s possible to obtain this amount of blood using accessories: Glucoject Dual S and Glucoject Lancets No-dol.

The blood glucose measurement starts only when the correct amount of blood is applied on the strip.

Sampling also from alternative sites

Results in 15 seconds

360 results are automatically stored in memory with date and time. Offering the possibility of calculating an average over 14 or 30 days. Results can be downloaded to your computer using the GlucoLog software.

 

 

Technical Data

Test

Glucose in whole blood

Sample volume

0,6 µL

Test strip

GLUCOCARD G sensor only

Measuring range

10-600 mg/dL

Measuring time

15 seconds

Environmental conditions

Temperature:10-40 °C; automatic compensation by using a built-in thermo sensor. Relative humidity: 20-80% (no dew condensation)

External output

a connection terminal for downloading data

Power source

3-Volt lithium battery (CR2032 o DL2032)

Battery life

Approximately 2000 tests (depending on the measurement conditions)

Low battery indication

Display battery symbol

Memory

360 test results

Dimensions

52 x 90 x 16 mm

Weight

Approximately 53 g (including battery).

   

GLUCOCARD G sensor

 

Glucocard G sensor

Meter

Only with GLUCOCARD G + and GLUCOCARD G meter

Measuring principle

Electrochemical.

Sample volume

0,6 µL

Response Time

5.5 seconds

Measuring range

10-600 mg/dL

Enzyme system

FAD-GDH (Flavin adenine dinucleotide - dehydrogenase).

Interference

Maltose and galactose free.The interferences of paracetamol, ascorbic acid, uric acid are reduced.

Expiry

1 year and 6 months from data of production. The expiy date is six months from the date of first opening the vial.

Accessories

Glucoject Dual S and Glucoject lancets No-Dol

Glucoject_DualS_penna

Glucoject Dual S Features

Comfort Zone Technology

Making blood sampling much less of a pain!

Damage or injury to the skin causes receptors to generate pain signals that are transmitted to the brain via spinal cord.

Through the particular Comfort Zone Technology (CZT), Glucoject Dual S uses eight raised points to apply pressure on nerve fibres in the finger and send a message of comfort to the brain whilst masking the message of pain.

According to the "gate theory ", the tactile stimulus to the brain, conveyed by large nervous fibres, obstructs the passage of the pain stimulus which is instead conveyed by smaller nervous fibres.

Glucocard accessories

In this way, Glucoject Dual S thanks to eight raised points causes a reduced perception of pain, when the lancet needle penetrates the skin for taking a blood sample.
Thanks to CZT, Glucoject Dual S increases comfort, masking the sensation perception of pain, making blood sampling much less of a pain.

21 Possibilities to customize the fingertip

- 3 force adjustment features
- 7 penetration depth settings (0 minimum depth, 6 maximum depth).

AST cap
To be used for alternative Sites

Glucocard fingertip

Needles fingertip: Glucoject Lancets No-Dol

  • Diameter tip: 30G (0,315 mm)
  • Tip length: 3,3 mm
  • Geometry needle: three-blade facets
  • Sterilization: gamma radiation
  • Use: disposable
  • Material: steel, latex free
  • Processing phases: Sharpening-Polishing,Sealing-Sterilisation

Glucoject Lancets No-Dol are compatible with most lancing systems on the market.

Glucoject Dual S can only be used with Glucoject Lancets No-Dol