Available in: Italy, Spain and Portugal
|
|
|
|
|
|
Features
|
|
|
No calibration is required
|
|
|
It requires 0,6 µL blood. It’s possible to obtain this amount of blood using accessories: Glucoject Dual S and Glucoject Lancets No-dol.
|
|
|
The blood glucose measurement starts only when the correct amount of blood is applied on the strip.
|
|
|
Sampling also from alternative sites
|
|
|
Results in 15 seconds
|
|
|
360 results are automatically stored in memory with date and time. Offering the possibility of calculating an average over 14 or 30 days. Results can be downloaded to your computer using the GlucoLog software.
|
|
|
|
|
|
|
Technical Data
|
|
|
Test
|
Glucose in whole blood
|
|
Sample volume
|
0,6 µL
|
|
Test strip
|
GLUCOCARD G sensor only
|
|
Measuring range
|
10-600 mg/dL
|
|
Measuring time
|
15 seconds
|
|
Environmental conditions
|
Temperature:10-40 °C; automatic compensation by using a built-in thermo sensor. Relative humidity: 20-80% (no dew condensation)
|
|
External output
|
a connection terminal for downloading data
|
|
Power source
|
3-Volt lithium battery (CR2032 o DL2032)
|
|
Battery life
|
Approximately 2000 tests (depending on the measurement conditions)
|
 |
|
Low battery indication
|
Display battery symbol
|
|
Memory
|
360 test results
|
|
Dimensions
|
52 x 90 x 16 mm
|
|
Weight
|
Approximately 53 g (including battery).
|
|
|
|
|
GLUCOCARD G sensor
|
|
|
|
|
Meter
|
Only with GLUCOCARD G + and GLUCOCARD G meter
|
|
Measuring principle
|
Electrochemical.
|
|
Sample volume
|
0,6 µL
|
|
Response Time
|
5.5 seconds
|
|
Measuring range
|
10-600 mg/dL
|
|
Enzyme system
|
FAD-GDH (Flavin adenine dinucleotide - dehydrogenase).
|
|
Interference
|
Maltose and galactose free.The interferences of paracetamol, ascorbic acid, uric acid are reduced.
|
|
Expiry
|
1 year and 6 months from data of production. The expiy date is six months from the date of first opening the vial.
|
|
|
|
Accessories
|
|
Glucoject Dual S and Glucoject lancets No-Dol
|
|
|
Glucoject Dual S Features
Comfort Zone Technology
Making blood sampling much less of a pain!
Damage or injury to the skin causes receptors to generate pain signals that are transmitted to the brain via spinal cord.
Through the particular Comfort Zone Technology (CZT), Glucoject Dual S uses eight raised points to apply pressure on nerve fibres in the finger and send a message of comfort to the brain whilst masking the message of pain.
According to the "gate theory ", the tactile stimulus to the brain, conveyed by large nervous fibres, obstructs the passage of the pain stimulus which is instead conveyed by smaller nervous fibres.
|
|
|
In this way, Glucoject Dual S thanks to eight raised points causes a reduced perception of pain, when the lancet needle penetrates the skin for taking a blood sample. Thanks to CZT, Glucoject Dual S increases comfort, masking the sensation perception of pain, making blood sampling much less of a pain.
21 Possibilities to customize the fingertip
- 3 force adjustment features - 7 penetration depth settings (0 minimum depth, 6 maximum depth).
AST cap To be used for alternative Sites
|
|
|
Needles fingertip: Glucoject Lancets No-Dol
- Diameter tip: 30G (0,315 mm)
- Tip length: 3,3 mm
- Geometry needle: three-blade facets
- Sterilization: gamma radiation
- Use: disposable
- Material: steel, latex free
- Processing phases: Sharpening-Polishing,Sealing-Sterilisation
Glucoject Lancets No-Dol are compatible with most lancing systems on the market.
Glucoject Dual S can only be used with Glucoject Lancets No-Dol
|
|
|
|
|
|
|
|
|
...