Diagnostic International / Products / Point of Care Testing / Spotchem EZ
 
Contact us - Sitemap
English
International
Spotchem EZ

POC Biochemistry

 

SYSTEM DESCRIPTION

  • Small, bench top fully automated biochemistry analyser
  • Internal centrifuge reduces manual handling of samples
  • Seven different multi-test panels available
  • Can measure up to 9 parameters simultaneously on a sample.
  • Simple calibration by magnetic card
  • Comprehensive 21 parameter test menu
  • Results available in minutes for rapid follow-up & actions
  • Calibration Check QC material provides assurance of result accuracy and precision
  • Connects to NetCare data manager.

Testing at Point of Care for Better Patient Care

Spotchem EZ

 

Single test strips

Measurement Range

 

Albumin

10-60

g/l

Alkaline Phosphatase

50-1,500

U/l

ALT (GPT)

10-1,000

U/l

Amylase

10-800

U/l

AST (GOT)

10-1,000

U/l

Calcium

0.75-4.99

mmol/l

Total Cholesterol

1.29-10.34

mmol/l

CPK

50-2,000

U/l

Creatinine

27-3,536

µmol/l

Fructosamine

50-1,000

µmol/l

Gamma-GT

10-1500

U/l

Glucose

1.1-25.0

mmol/l

Direct HDL Cholesterol

0.26-3.88

mmol/l

LDH

100-4,000

U/l

Magnesium

0.08-2.47

mmol/l

Inorganic Phosphorus

0.16-6.46

mmol/l

Total Bilirubin

3-428

µmol/l

Total Protein

20-110

g/l

Triglyceride

0.28-5.65

mmol/l

BUN (Urea)

1.8-71.5

mmol/l

Uric Acid

59-1190

µmol/l

Multi-parameter strips

Parameters

Panel-V

Alk. Phosphatase, ALT, Creatinine, Glucose,
T. Protein, BUN

Kidney-2

Albumin, Creatinine, T. Protein, BUN, Uric Acid

Liver-1

Albumin, ALT, AST, LDH, T. Bilirubin, T. Protein

Heart-2

AST, CPK, T. Cholesterol, LDH, T. Protein, BUN

Panel-1

ALT, AST, T. Cholesterol, Glucose, T. Bilirubin, BUN

Panel-2

Albumin, Calcium, LDH, Triglyceride,
T. Protein, Uric Acid

Stat-1

ALT, AST, CPK, LDH, T. Bilirubin, BUN

TECHNICAL SPECIFICATION

Measurement principle:

Dual wavelength reflectometry with end-point and kinetic assays

Sample type:

Serum, Plasma, Whole Blood

Sample required:

7µl per test + 38µl dead volume: serum, plasma

Parameters available:

21

Simultaneous measurement:

1-9 parameters (3 single strips plus 1 multi strip)

Throughput:

63 tests/hour (9 tests per batch)

Calibration method:

Magnetic card

Memory:

100 tests

Display:

20 digits x 2 lines LCD

Built-in printer:

36-character thermal printer

External output:

RS-232C interface

Power supply:

AC100-120V, 220-240V, 50-60Hz

Dimensions:

338mm(W) x 203mm(D) x 167mm(H)

Weight:

Approximately 5.4kg

Operating temperature:

10-30°C