POC Biochemistry
|
|
|
SYSTEM DESCRIPTION
-
Small, bench top fully automated biochemistry analyser
-
Internal centrifuge reduces manual handling of samples
-
Seven different multi-test panels available
-
Can measure up to 9 parameters simultaneously on a sample.
-
Simple calibration by magnetic card
-
Comprehensive 21 parameter test menu
-
Results available in minutes for rapid follow-up & actions
-
Calibration Check QC material provides assurance of result accuracy and precision
-
Connects to NetCare data manager.
|
|
|
|
|
|
Single test strips
|
Measurement Range
|
|
|
Albumin
|
10-60
|
g/l
|
|
Alkaline Phosphatase
|
50-1,500
|
U/l
|
|
ALT (GPT)
|
10-1,000
|
U/l
|
|
Amylase
|
10-800
|
U/l
|
|
AST (GOT)
|
10-1,000
|
U/l
|
|
Calcium
|
0.75-4.99
|
mmol/l
|
|
Total Cholesterol
|
1.29-10.34
|
mmol/l
|
|
CPK
|
50-2,000
|
U/l
|
|
Creatinine
|
27-3,536
|
µmol/l
|
|
Fructosamine
|
50-1,000
|
µmol/l
|
|
Gamma-GT
|
10-1500
|
U/l
|
|
Glucose
|
1.1-25.0
|
mmol/l
|
|
Direct HDL Cholesterol
|
0.26-3.88
|
mmol/l
|
|
LDH
|
100-4,000
|
U/l
|
|
Magnesium
|
0.08-2.47
|
mmol/l
|
|
Inorganic Phosphorus
|
0.16-6.46
|
mmol/l
|
|
Total Bilirubin
|
3-428
|
µmol/l
|
|
Total Protein
|
20-110
|
g/l
|
|
Triglyceride
|
0.28-5.65
|
mmol/l
|
 |
|
BUN (Urea)
|
1.8-71.5
|
mmol/l
|
|
Uric Acid
|
59-1190
|
µmol/l
|
|
Multi-parameter strips
|
Parameters
|
|
Panel-V
|
Alk. Phosphatase, ALT, Creatinine, Glucose, T. Protein, BUN
|
|
Kidney-2
|
Albumin, Creatinine, T. Protein, BUN, Uric Acid
|
|
Liver-1
|
Albumin, ALT, AST, LDH, T. Bilirubin, T. Protein
|
|
Heart-2
|
AST, CPK, T. Cholesterol, LDH, T. Protein, BUN
|
|
Panel-1
|
ALT, AST, T. Cholesterol, Glucose, T. Bilirubin, BUN
|
|
Panel-2
|
Albumin, Calcium, LDH, Triglyceride, T. Protein, Uric Acid
|
|
Stat-1
|
ALT, AST, CPK, LDH, T. Bilirubin, BUN
|
|
|
|
TECHNICAL SPECIFICATION
|
|
Measurement principle:
|
Dual wavelength reflectometry with end-point and kinetic assays
|
|
Sample type:
|
Serum, Plasma, Whole Blood
|
|
Sample required:
|
7µl per test + 38µl dead volume: serum, plasma
|
|
Parameters available:
|
21
|
|
Simultaneous measurement:
|
1-9 parameters (3 single strips plus 1 multi strip)
|
|
Throughput:
|
63 tests/hour (9 tests per batch)
|
|
Calibration method:
|
Magnetic card
|
|
Memory:
|
100 tests
|
|
Display:
|
20 digits x 2 lines LCD
|
|
Built-in printer:
|
36-character thermal printer
|
|
External output:
|
RS-232C interface
|
|
Power supply:
|
AC100-120V, 220-240V, 50-60Hz
|
|
Dimensions:
|
338mm(W) x 203mm(D) x 167mm(H)
|
|
Weight:
|
Approximately 5.4kg
|
|
Operating temperature:
|