Test strip
GlucoMen LX sensor
Sample
Fresh capillary whole blood
Sample volume
0.3 µL
Measuring principle
Electrochemical
Enzyme system
Glucose oxidase
Testing time
4 seconds
Test range
20-600 mg/dL (1.1-33.3 mmol/L)
Testing conditions
Temperature: 5-45 °C (41-113 °F)
Relative humidity: 10-90% (no dew condensation)
Icodextrin interference
The GlucoMen LX PLUS system is suitable for use by
people on peritoneal dialysis using solutions containing
icodextrin.
Device standard
EN ISO 15197:2015
Batteries
Two 3.0 V batteries (CR2450)
Battery life
Approx. 1000 tests
Memory
400 test results
Averages
Over 1, 7, 14, 30 days
Dimensions
58 x 98 x 17 mm
Weight
75 g
Data transmission
To PC (via USB cable)
To mobile devices (via external Bluetooth)
β-KETONE MEASUREMENT SPECIFICATIONS
Test strip
GlucoMen LX ß-Ketone sensor
Sample volume
0.8 µL
Measuring principle
Electrochemical
Enzyme system
ß-Hydroxybutyrate dehydrogenase
Testing time
10 seconds
Test range
0,1-8,0 mmol/L
Testing conditions
Temperature: 5-45 °C (41-113 °F)
Relative humidity: 10-90%