Dimension
74 cm H, 54 cm W, 66 cm D |
Weight
62 Kg |
Element of analysis
Quantification of glycated haemoglobin |
Principle
Boronate affinity high performance liquid chromatography |
Minimum readable division
Area percent = 0,01% |
Minimum sample requirements
10 μL whole blood, 5 μL packed red blood cells |
Dilution
Whole blood 1 : 150, packed red blood cells 1 : 300 |
Sample capacity
210 samples + STAT position |
Throughput
75 seconds per injection |
Analytical column
Boronate bonded to a porous polymer support gel |
Column temperature
55 °C |
Detection
LED wavelength detector, 413 nm |
Software
lnstalled on hard drive of Hb9210 Premier CPU |
Calibrators
On board, Glycated Haemoglobin Calibrators (2 levels) |
Controls