HA-8180T

The information to be found in this section is intended for the exclusive use of healthcare professionals.

Samples:
Whole blood or hemolysis samples
Measurement items:
HbA1c, HbA2 and HbF (HbS, HbC, HbD, HbE can be detected)
Measurement principle:
Reversed-phase cation exchange liquid chromatography
Measurement method:
Dual wavelength colorimetry (420/500 nm)
Measurement range:
HbA1c,: 3-20%, 9-195 mmol/mol; HbA2: 2-10%; HbF: 0-100%
Resolution:
0.1%, 1 mmol/mol
Processing speed:
210 seconds per sample
Sample container:
Tube: 12.3-15 mm diameter; 75-100 mm length; Cup: 500 μl
Compatible rack type:
ARKRAY racks
Sample consumption:
14 μl (whole blood)
Required sample volume:
Tube: minimum 10 mm from the bottom; Cup: 400 μl
Sampling method:
Sample tube stirring and cap piercing
Sample loading capacity:
Maximum 100 samples, continuous loading
Column temperature:
Approximately 39°C
Display:
Large color LCD with backlight
Built-in printer:
Thermal printer (58 mm thermal paper
Memory capacity:
900 measurement results (including calibration results
Operating environment:
Temperature: 10-30°C. Humidity: 20-80% no condensation
External output:
1 serial port (can be used as Ethernet port)
Communication system:
RS-232C compliant (one-way or two-way)
Power requirements:
AC100-240V ± 10%, 50/60 Hz
Power input:
Maximum 300 VA
Dimensions:
530 (W) x 530 (D) x 530 (H) mm
Weight:
Analyzer: 39 kg; Sampler: approx. 4 kg